- Dimensions: 400 x 330 x 101 mm, approx. 4.8 kg
- Power supply requirements: 220 - 240 V (nominal), 50/60 Hz or 100 - 115 V (nominal), 50/60 Hz independent operation with built-in rechargeable battery, LED indication for mains and battery, built-in charger unit
- Power consumption: Max. 28 VA
- Battery charging time: 15 hrs. from a completely discharged battery, 3 hrs. to 60 % capacity
- Battery capacity: 6 hrs. of normal use or 400 automatic ECG printouts Frequency range of digital recorder: 0 Hz - 150 Hz (IEC/AHA)
- Patient input circuitry: Fully floating and isolated, defibrillation protected (only with original SCHILLER patient cable)
- Control panel and keyboard: User-friendly key pad, LED indicators
- LED indicators: > Lead group > Mains connected > Battery operation and low battery (indicator flashes) > Paper tray empty or jammed > Filter status (on/off) > Loose electrodes
- Leads: 12 simultaneous leads: Standard / Cabrera Myogram (muscle tremor) filter: Adjustable at 25 or 35 Hz (also for copy function)
- Line frequency filter: Distortion-free suppression of superimposed 50 or 60 Hz sinusoidal interferences by means of adaptive digital filtering ECG amplifier: > Simultaneous, synchronous recording of all 9 active electrode signals (= 12 leads) > Sampling frequency: 1000 Hz > Pacemaker detection: Amplitudes from 2 to 700 mV and pulse widths from 0.1 to 2.0 ms
|
|
|
 |
|
| |
|